Powerheart AED G3Pro 9300P-501

Daily Maintenance
Check the Status Indicator to ensure that it is GREEN. When the indicator is GREEN, the AED is ready for a rescue. If the indicator is RED, refer to the Troubleshooting Table in Operations Manual





Monthly Maintenance
Perform the following procedure each month (28 days)
1. Open the AED lid.
2. Wait for the AED to indicate status: Observe the change of the STATUS INDICATOR to RED. After approximately 5 seconds, verify that the STATUS INDICATOR returns to GREEN. 
3. Check the expiration date on the electrodes.
4. Listen for the voice prompts.
The battery power is completely depleted. Replace with a new battery. If STATUS INDICATOR remains RED inform SEDAFIAT Technican.
5. Close the lid and observe the change of the STATUS INDICATOR to RED. After approximately 5 seconds, verify that the STATUS INDICATOR returns to GREEN.

Document the inspection on an AED inspection tag or separate document.



Annual Maintenance
Perform the following tests annually to confirm that the diagnostics are functioning properly and to verify the integrity of the case.
Check the Integrity of the Pads and Circuitry:
1. Open the AED lid.
2. Remove the pads.
3. Close the lid.
4. Confirm that the STATUS INDICATOR turns RED.
5. Open the lid and confirm that the Pad indicator is lit.
6. Reconnect the pads and close the lid.
7. Make sure the expiration date is visible through the clear window of the lid. 
8. Check to make sure that the STATUS INDICATOR is GREEN. If the pads are not installed properly, the PAD indicator will illuminate; call Customer Service for assistance.
9. Open the lid and confirm that no diagnostic indicators are lit.
10. Check the expiration date of the pads; if expired, replace them.
11. Check the pads packaging integrity.
12. Close the lid Check the Integrity of the Service Indicator (LED) and Circuitry:
1. Immediately after opening the AED lid, press and hold the Shock button and confirm that the Service LED is lit.
2. Release the Shock/Continue button.
3. Close the lid.
4. Verify that the STATUS INDICATOR remains RED.
5. Open the lid and confirm that no diagnostic indicators are lit.
6. Close the lid.
7. Verify that the STATUS INDICATOR turns GREEN. 

CAUTION: Case Cleaning Solutions. When disinfecting the case, use a non-oxidizing disinfectant, such as ammonium salts or a glutaraldehyde based cleaning solution, to avoid damage to the metal connectors.
 

For more Information on Cardiac Science AED Maintenance visit our Cardiac Science Resources Pages